Pressure Testing & Optical Measurement


We’re ready to help you make sure your subsea instrumentation is ready to perform under pressure. We have multiple hydrostatic pressure chambers available to meet a broad range of requirements and experienced technicians on staff to implement your testing plan.

Please fill out and submit the Pressure Test Order Form at least two weeks prior to your desired testing date. A member of our Sales team will review your request and reach out to you with an estimate.

Need to cancel or reschedule? Read our cancellation and rescheduling policy.


  • Operating pressures up to 30,000 psi
  • Capacity up to 508 mm (20 in) diameter and 1219.2 mm (48 in) in height
  • Feed-through options and adapters available
  • Annually certified gauges ensure accuracy across all pressure ranges
  • Long-duration holds and repeated cyclical testing capable

6-in. 6×24/11L20-in. 20×48/247L
ShapeCylindricalCylindrical
Interior Diameter152 mm (6 in)508 mm (20 in)
Interior Height609 mm (24 in)up to 1,219 mm (48 in)
Max Pressure206.5 MPa (30,000 psi)137.5 MPa (20,000 psi)
Max Ramp Up Rate3,000 psi/min800 psi/min (approx.)
Max Ramp Down Rate5,000 psi/min5,000 psi/min
Feed-thru Options Adapters Available1 or 2
3/4″ – 16 threads
1 or 2
1″ – 14 threads

Cancellation and Rescheduling Policy

Notice GivenCancellation FeeRescheduling Fee
7+ calendar daysNo chargeNo charge
4-6 calendar days25% of quoted value$500 per instance
2-3 calendar days50% of quoted value$500 per instance
<24 hours100% of quoted value100% of quoted value

We use our in-house integrating sphere to measure light output across the entire beam pattern and to validate LED output and coloration as part of our design and manufacturing process. The 1 meter sphere is available to test lights from customers and other manufacturers. We ship every LED SeaLite, SeaLite Sphere, SeaLite Lumos Remote, and Nano SeaLite with its own True Lumens Report and the confidence that you will get the specified output and coloration.


  • 1 meter diameter sphere
  • Integrated spectrometer
  • Multiple luminary positions
  • Lumens, CRI, and spectral output


An integrating sphere, also known as an Ulbricht or integrated spectrometer sphere, is used to concisely capture the flux (total power) of a light source as well as precisely measure the color of the light. Utilizing the concept of Lambertian reflectance to scatter the light in a diffuse manner, the total light energy created is measured without inaccuracies caused by directional characteristics of the lights’ source.

The inside of the sphere has a non-fluorescent reflective white coating which effectively scatters and distributes the light equally to all other points. We use this to measure the total light from the lamp over all angles of illumination and observation.

With a readout in Lumens given and a mapping of color temperature against a black body graph, a complete characteristic of the light is quantified. A True Lumens Report will be generated and provided for each light that is tested. The sphere’s light-port accepts apertures up to a maximum diameter of 3.5 in. with adapters available for smaller dimensions.